ABBOTT GMBH MAGNESIUM; PHOTOMETRIC METHOD, MAGNESIUM Back to Search Results
Model Number 03P6822
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/14/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The search by lot number did not find elevated complaint activity.The trend review by the product list number found no adverse trends related to this issue.Labeling is adequate and provides specific instructions to the customer regarding the current issue.Review in the corrective and preventive actions system did not identify any non-conformances or deviations related to the to the likely cause list number/lot number and complaint issue.As part of troubleshooting the sample was retested and results were within range.No additional issues were identified.No product deficiency was identified.
 
Event or Problem Description
The account generated false elevated architect magnesium of (b)(6) mg/dl on a sample ((b)(6) year old, female, caucasian patient) that repeated (b)(6) mg/dl.The account uses a normal magnesium range of 1.6 to 2.6 mg/dl.No impact to patient management was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGNESIUM
Common Device Name
PHOTOMETRIC METHOD, MAGNESIUM
Manufacturer (Section D)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM  65205
Manufacturer (Section G)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM   65205
Manufacturer Contact
siobhan wright
lisnamuck
post market surveillance
longford N39 E-932
433331157
MDR Report Key13477124
Report Number3002809144-2022-00046
Device Sequence Number17515076
Product Code JGJ
UDI-Device Identifier00380740161620
UDI-Public00380740161620
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K181748
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 02/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/17/2022
Device Model Number03P6822
Device Catalogue Number03P68-22
Device Lot Number09150UN21
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/14/2022
Initial Report FDA Received Date02/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/28/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARC C4000 INTGR, 02P24-40, C401897
Patient Age79 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-