• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON LABORATORIES IRELAND LTD. CLAREON IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON LABORATORIES IRELAND LTD. CLAREON IOL; INTRAOCULAR LENS Back to Search Results
Model Number SY60WF
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/12/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was not returned for analysis.Video was provided.Video provided shows iol implantation.Type of accessories used with iol implantation such as cartridge and handpiece cannot be confirmed from the video.The preparation of the product is not recorded.The lens position at the pause location cannot be confirmed, but from available footage it appears to be acceptable.Viscoelastic (ovd) line is visible in the cartridge tip, no exact determination on the ovd fill can be made, however it appears that the cartridge may not have been fully filled with ovd as per instructions for use (ifu) instruction.The lens is implanted, trailing haptic is observe to be attached to the optic for a few second, it releases with no issues.There appears to be marks/lines over the optic edge near the trailing haptic ( at around 7 o'clock position).The lens is cut in half and removed.The complainant indicates non company viscoelastic, which is not qualified for use with the associated iol model.The product investigation could not identify the root cause for the reported complaint "crack was found in iol optic".The product and implantation accessories such as cartridge was not returned for evaluation.Video evaluation confirmed mark on the optic edge.The root cause may be related to failure to follow ifu, as the customer states the use of non-qualified viscoelastic.The use of an unqualified ovd may cause damage to the lens and potential complications during the implantation process.The ifu instructs: fill the device until ovd can be observed flowing to the nozzle tip.This will require approximately 0.28 ml of viscoelastic.If inadequate viscoelastic is placed in the device this will cause inadequate coverage of the lens fold path which may cause the lens to advance incorrectly and cause damage.Lens damage may occur: due to the use of a non-qualified viscoelastic.Lens delivery performance may be negatively affected when using other non-qualified viscoelastics leading to underfill, overfill, misfolding, delivery issues and /or damage.If inadequate viscoelastic is placed in the device this will cause inadequate coverage of the lens fold path which may cause the lens to advance incorrectly and cause delivery issues and /or damage.If the operating room temperature is too high (> 23°c / 73° f) lens folding consistency is negatively affected, the lens is more adherent and this may inhibit lens advancement and cause delivery issues and / or product damage.Any of the above listed causes alone, or in combination, may create the reported event.In addition to this, all iols are 100% cosmetically inspected as per approved manufacturing procedures and the observed lens damage would not meet our current release criteria.Based on these investigation findings, we are unable to verify if the iol/device contributed to the event.All product and batch history records are quality reviewed prior to product release.There have been no other complaints for this lot.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A physician reported that, during intraocular lens (iol) implant procedure, the optic was found cracked after insertion of iol.The surgery was performed and completed after replacing the product with another lens.Additional information was requested.
 
Additional Manufacturer Narrative
The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
Additional information was obtained from the surgeon that, it took about 1 minute on average from setting iol to the cartridge to inserting it.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI  00000
Manufacturer (Section G)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI   00000
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key13477582
Report Number9612169-2022-00041
Device Sequence Number20136329
Product Code HQL
UDI-Device Identifier00380652251365
UDI-Public00380652251365
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2023
Device Model NumberSY60WF
Device Lot Number21280377
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/12/2022
Supplement Date Received by Manufacturer02/07/2022
Initial Report FDA Received Date02/08/2022
Supplement Report FDA Received Date03/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/01/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CLAREONINJECTOR; FORCEPS; MONARCH III IOL CARTRIDGE D; OPELEAD
Patient SexUnknown
-
-