• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON LABORATORIES IRELAND LTD. ACRYSOF SINGLEPIECE IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON LABORATORIES IRELAND LTD. ACRYSOF SINGLEPIECE IOL; INTRAOCULAR LENS Back to Search Results
Model Number SA60AT
Medical Device Problem Codes Material Twisted/Bent (2981); Scratched Material (3020)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/05/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A nurse reported that, during intraocular lens (iol) implant procedure, while deploying the lens in sac it is observed that it is deteriorated in the optics with 2 lines, apparently perpendicular to haptic peripheral fractures, one of them partially composes the pupillary axis.The lens shifted.Additional information was received that, the procedure was completed without additional complications.Lens extraction was performed and replaced with a new one.
 
Additional Manufacturer Narrative
The product was returned for analysis and the reported complaint was observed.Iol returned adhered inside gauze tape.Optic can be seen to be cut.Unable to remove sample from gauze tape or conduct further evaluation testing due to condition of returned sample.The complainant indicates the use of non-company as a viscoelastic, which is not qualified for use with associated iol model.While we are unable to determine the origin of the reported complaint, our observations reasonably suggest that it is not manufacturing related.Based on the information provided by the customer, the most likely root cause is failure to follow instructions for use (ifu), as the surgeon states the use of non-qualified combination.The use of an unqualified combination may cause damage to the iol and potential complications during the implantation process.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF SINGLEPIECE IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI  00000
Manufacturer (Section G)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI   00000
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key13478072
Report Number9612169-2022-00044
Device Sequence Number20108968
Product Code HQL
UDI-Device Identifier00380652555876
UDI-Public00380652555876
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 05/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/06/2026
Device Model NumberSA60AT
Device Catalogue NumberSA60AT.235
Device Lot Number25160556
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/13/2022
Supplement Date Received by Manufacturer04/26/2022
Initial Report FDA Received Date02/08/2022
Supplement Report FDA Received Date05/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/08/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MONARCH III IOL CARTRIDGE D; MONARCH IOL INJECTOR; OVD: BIOVISC
Patient Age81 YR
Patient SexFemale
-
-