• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER TMT TRABECULAR METAL ACETABULAR REVISION SYSTEM ACETABULAR AUGMENT; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER TMT TRABECULAR METAL ACETABULAR REVISION SYSTEM ACETABULAR AUGMENT; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Code Failure of Implant (1924)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the device remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Additional associated products and mdrs 110010267 g7 osseoti multihole 58mm g lot# 7084562; mdr: 0001825034-2022-00232.
 
Event or Problem Description
It was reported that the patient had a total hip arthroplasty.The patient was non-compliant (due to the patient not following restrictions at rehab) and the cup, augment block, and screws loosened.Therefore, the patient was revised with a bigger cup, bigger augment, and screws approximately 1-2 months later.Subsequently, the patient was non-compliant again, therefore will be revised again in the near future with a custom trifflange.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.No product was returned; visual and dimensional evaluations could not be performed.Scanned and electronic device history records for were reviewed for deviations and/ or anomalies.Records identified no deviations or anomalies during manufacturing related to the reported event.The reported products were reviewed for compatibility with no issues noted.Review of complaint history found no additional related issues for this item and the reported part and lot combination.Medical records were not provided.The root cause of the reported issue is attributed to the patient being non-compliant to post operative care as reported by the surgeon.Per ifu/package insert under the important notes section 'zimmer biomet is not liable for complications arising from the use of the device outside of its indicated uses, surgical technique or judgment, product selection, and similar matters outside the control of zimmer biomet.Complications and/or failure of prosthetic implants are more likely to occur in patients with unrealistic functional expectations, heavy patients, physically active patients, and/or with patients who fail to follow through with the required rehabilitation program.Physical activity can result in loosening, wear, and/or fracture of the implant.The prospective implant patient must be counseled about the capabilities of the implant and the impact it will have on his or her lifestyle.¿ no corrective actions, preventive actions, or field actions resulted after investigation of this event.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRABECULAR METAL ACETABULAR REVISION SYSTEM ACETABULAR AUGMENT
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER TMT
10 pomeroy road
parsippany NJ 07054
Manufacturer (Section G)
ZIMMER TMT
10 pomeroy road
parsippany NJ 07054
Manufacturer Contact
chrsitina arnt
10 pomeroy road
parsippany, NJ 07054
9735760032
MDR Report Key13479282
Report Number3005751028-2022-00001
Device Sequence Number5211435
Product Code LPH
UDI-Device Identifier00889024306783
UDI-Public(01)00889024306783(17)250531(10)64634283
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K042871
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2025
Device Model NumberN/A
Device Catalogue Number00489405020
Device Lot Number64634283
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/12/2022
Supplement Date Received by Manufacturer04/07/2022
Initial Report FDA Received Date02/08/2022
Supplement Report FDA Received Date04/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/29/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexFemale
-
-