Model Number 71303600
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 01/20/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, after a thr had been performed on (b)(6) 2020, the patient experienced recurrent dislocations.The hip dislocated on 5 separate occasions.Therefore, a revision surgery was performed on (b)(6) 2022 in order to solve this adverse event.It was found that the hip was relatively stable, and the cocr 12/14 fem head 36mm +0 was replaced for a longer femoral head component which increased the stability significantly.The current health status of patient is unknown.
 
Additional Manufacturer Narrative
Internal complaint reference: case (b)(4).
 
Additional Manufacturer Narrative
H3, h6: the device was not returned for evaluation.The pictures provided were reviewed and could not confirm the stated failure mode.The clinical/medical investigation concluded that, the provided x-ray was reviewed and does not contribute to the medical investigation, although the fact of longer neck during revision indicates laxity within the joint that was improved.Smith and nephew has not received the device/ adequate materials to fully evaluate the complaint, but if additional clinically relevant materials are later received, then the case may be re-opened for further evaluation.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the instructions for use documents for total hip systems revealed that dislocation has been identified as leading complication associated with revision surgery.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible causes could include but not limited to traumatic injury, size of the device, patient anatomy or abnormal loading of limb.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COCR 12/14 FEM HEAD 36MM +0
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13479846
Report Number1020279-2022-00554
Device Sequence Number14723767
Product Code LPH
UDI-Device Identifier03596010479983
UDI-Public03596010479983
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K022902
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71303600
Device Catalogue Number71303600
Device Lot Number20CM20716
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/20/2022
Supplement Date Received by Manufacturer03/08/2022
Initial Report FDA Received Date02/08/2022
Supplement Report FDA Received Date03/13/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/25/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexUnknown
-
-