H10 h3, h6 part of the reported device was received for evaluation.There was no way to determine if the device contributed to the reported event.A complaint history review found no similar reported events.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the customer provided image found a deployed anchor next to a bent shaft.No damage to the anchor can be seen in the image.A visual inspection of the returned device found that it is not in its original packaging.The device has been deployed, and no anchor or sutures were received.The shaft of the device is bent, and there is debris on the device.A functional evaluation could not be performed due to the condition in which the device was received.Per case details, the anchor was pulled out from the patient along with the needle.The procedure was completed using a back-up device.No further risk to the patient is anticipated as a result of the additional bone hole.Since no further harm is anticipated no further clinical assessment is warranted at this time.The complaint was not confirmed, and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.Factors that could have contributed to the reported event include inadequate bone quality, improper preparation of the insertion site, or off-axis insertion of the device.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
|