| Model Number |
637-202 |
| Medical Device Problem Code |
Excess Flow or Over-Infusion (1311)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical inc.Internal report number (b)(4).The complaint is under evaluation.A follow-up report will be provided after the examination results are available.
|
| |
|
Event or Problem Description
|
|
As reported by the user facility: "infusions are going in twice as fast as they should.Literally a 4 min infusion finished in 2 minutes with the patients getting extreme side effects." no additional information provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical inc.Internal report number (b)(4).Failure analysis and investigation results did not confirm the reported issue.Upon receipt the actual pump involved was visually and functionally inspected.Three (3) volumetrics of 250ml/hr and 25ml were performed and tested in spec.The pump operated as intended and the reported failure could not be reproduced.Based on the results of the investigation, no conclusions can be made regarding the cause of the reported event.If additional information becomes available, a follow up report will be submitted.
|
| |
|