Model Number 142122-11
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Restenosis (4576)
Date of Event 01/28/2022
Type of Reportable Event Serious Injury
Event or Problem Description
This mdr is related to mdr # 30116321500-2022-00006.There is no evidence to suggest that the device caused or contributed to this event.The patient was treated as part of the (b)(6) study on (b)(6) 2021.At index ,the subject had two biomimics stents implanted, one in the sfa middle third of the left leg and the other from the sfa distal third to proximal popliteal of the left leg.On (b)(6) 2022, a restenosis of treated segment was identified in the sfa proximal third to sfa distal third and treated with percutaneous transluminal angioplasty / standard balloon angioplasty, laser/atherectomy, and thrombectomy.The event is reported as "definitely related" to the device and target lesion related.The event is resolved and patient has recovered.The device remains implanted.
 
Additional Manufacturer Narrative
This mdr is related to mdr # 30116321500-2022-00006.There was no reported device malfunction and the device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Based on the case information and related record review, a conclusive cause for the reported patient effects, and the relationship to the product, if any, cannot be determined.The reported patient effect of restenosis leading to additional surgical intervention is listed in the biomimics 3d instructions for use and is a known patient effect of peripheral stenting procedures.There is no indication of a product quality issue with respect to the design, manufacture or labelling of the device.If further information regarding this event becomes available, a follow-up report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOMIMICS 3D VASCULAR STENT SYSTEM
Common Device Name
BIOMIMICS 3D VASCULAR STENT SYSTEM
Manufacturer (Section D)
VERYAN MEDICAL LTD
unit 11 galway technology park
parkmore road
galway city, H91 V E0H
EI  H91 VE0H
Manufacturer Contact
alan mcdonagh
unit 11 galway technology park
parkmore road
galway, H91 V-E0H
EI   H91 VE0H
MDR Report Key13480320
Report Number3011632150-2022-00007
Device Sequence Number5462732
Product Code NIP
UDI-Device Identifier05391526850466
UDI-Public(01)05391526850466(17)220702(11)201201(10)0000061256
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P180003
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Study,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/02/2022
Device Model Number142122-11
Device Catalogue Number142122-11
Device Lot Number0000061256
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/28/2022
Initial Report FDA Received Date02/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/01/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ASPIRIN; CLOPIDOGREL (PLAVIX)
Outcome Attributed to Adverse Event Required Intervention;
Patient Age76 YR
Patient SexMale
Patient Weight79 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-