• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AXIOS STENT AND ELECTROCAUTERY ENHANCED DELIVERY SYSTEM; PANCREATIC STENT, COVERED, METALLIC, REMOVABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AXIOS STENT AND ELECTROCAUTERY ENHANCED DELIVERY SYSTEM; PANCREATIC STENT, COVERED, METALLIC, REMOVABLE Back to Search Results
Model Number M00553670
Medical Device Problem Code Activation, Positioning or Separation Problem (2906)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Approximated based on the date the manufacturer became aware of the event.(b)(4).The device has been received for analysis; however, the analysis has not yet been completed.Upon completion of the problem analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation on (b)(6) 2022 that a hot axios stent was to be implanted transgastric to the pancreas to treat a pseudocyst during an endoscopic ultrasound (eus) procedure performed on an unknown date.During the procedure, the second flange could not be deployed.The stent was removed partially deployed from the patient and the procedure was completed using another axios stent.There were no patient complications as a result of this event.
 
Additional Manufacturer Narrative
Block b3: approximated based on the date the manufacturer became aware of the event.Block h6: medical device problem code a15 captures the reportable event of stent partially deployed.Block h10: a hot axios stent and delivery system were returned for analysis.Visual examination of the returned device found the stent partially deployed.The delivery system was received in position "1".A jagwire was also returned stuck in the device.Microscopic inspection was performed and friction marks were noted on the handle and the sheath.X-ray analysis was performed to analyze the distal section of the sheath and the luer section; no damage was found.Functional inspection was performed and the stent could be released from the delivery system.No other issues were noted to the stent and delivery system.The reported event of stent partially deployed was confirmed via visual inspection.Taking all available information into consideration, the investigation concluded that the reported event and the observed failures were likely due to factors encountered during the procedure.It may be that how the device was handled and manipulated during the procedure, limited the performance of the device and contributed to the stent partial deployment and the guidewire being stuck in the catheter.Additionally, continuous attempts to deploy the stent could have led to the friction marks on the sheath and the handle.Therefore, a review and analysis of all available information indicated the most probable cause is adverse event related to procedure.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.
 
Event or Problem Description
It was reported to boston scientific corporation on (b)(6) 2022 that a hot axios stent was to be implanted transgastric to the pancreas to treat a pseudocyst during an endoscopic ultrasound (eus) procedure performed on an unknown date.During the procedure, the second flange could not be deployed.The stent was removed partially deployed from the patient and the procedure was completed using another axios stent.There were no patient complications as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AXIOS STENT AND ELECTROCAUTERY ENHANCED DELIVERY SYSTEM
Common Device Name
PANCREATIC STENT, COVERED, METALLIC, REMOVABLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key13480376
Report Number3005099803-2022-00555
Device Sequence Number5387570
Product Code PCU
UDI-Device Identifier08714729979333
UDI-Public08714729979333
Combination Product (Y/N)N
Initial Reporter StateSD
Initial Reporter CountryUS
PMA/510(K) Number
K192043
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/09/2023
Device Model NumberM00553670
Device Catalogue Number5367
Device Lot Number0027801393
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/01/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/17/2022
Supplement Date Received by Manufacturer02/10/2022
Initial Report FDA Received Date02/08/2022
Supplement Report FDA Received Date03/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/09/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-