| Model Number |
ETRIO335H |
| Medical Device Problem Code |
Difficult to Open or Close (2921)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/18/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during an unknown procedure, just out of the box, the jaws wouldn¿t open and close.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Batch #: v96d11.Additional information was requested and the following was obtained: did the patient suffer any adverse consequences? no; the nurse noticed that the jaws weren¿t working before the instrument made it close to the patient.They simply opened a new instrument and proceeded to the surgery.Delayed the surgery less than a few minutes.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device and the manufacturing criteria were met prior to the release of this lot/batch.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 3/7/2022.Investigation summary: the product was returned to ethicon endo surgery for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the etrio335h device was received with no apparent damage.The device was tested on the generator and passed all functional testing.The energy output delivered from the device was verified.All three tones were heard during functional testing (tone 1 is heard when the energy activation button is pressed; tone 2 is heard when tissue impedance threshold is reached and tone 3 is heard when the cycle is complete).No issues were noted with the opening and closing of the jaws.There were no anomalies noted with the functionality of the device.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.Although no conclusion could be reach on the cause of the reported event.The event described could not be confirmed as the device performed without any difficulties noted.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post market surveillance.
|
| |
|
Search Alerts/Recalls
|