• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. BIVONA; TUBE TRACHEOSTOMY AND TUBE CUFF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. BIVONA; TUBE TRACHEOSTOMY AND TUBE CUFF Back to Search Results
Model Number 67SP050
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/16/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Two tracheostomy tubes were received for evaluation in relation to the reported event.Visual inspection of the samples was able to confirm the reported event.Both samples were noted to be broken in two as described by the reported event.Root cause was unable to be determined, as it is not possible to exclude the potential for mishandling by the user in adjusting the tracheostomy as a cause.A dhr review was performed subsequent to the manufacturing of the device and prior to its release.No problems or issues were identified during this dhr review.
 
Event or Problem Description
It was reported that one of the patients trach tubes broke at the shaft.There was no patient injury reported.
 
Additional Manufacturer Narrative
Two tracheostomy tubes were received for evaluation in relation to the reported event.Visual inspection of the samples was able to confirm the reported event.Both samples were noted to be broken in two as described by the reported event.Root cause was unable to be determined, as it is not possible to exclude the potential for mishandling by the user in adjusting the tracheostomy as a cause.A dhr review was performed subsequent to the manufacturing of the device and prior to its release.No problems or issues were identified during this dhr review.
 
Event or Problem Description
It was reported that one of the patients trach tubes broke at the shaft.There was no patient injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIVONA
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
5700 west 23rd ave
gary IN 46406
Manufacturer (Section G)
SMITHS MEDICAL ASD, INC.
5700 west 23rd ave
gary IN 46406
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key13480714
Report Number3012307300-2022-02884
Device Sequence Number20170694
Product Code JOH
UDI-Device Identifier15021312000740
UDI-Public15021312000740
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K083641
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 02/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date12/03/2025
Device Model Number67SP050
Device Lot Number4086009
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/30/2021
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/20/2021
Initial Report FDA Received Date02/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/02/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age7 YR
Patient SexMale
-
-