• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BLUE BELT TECHNOLOGIES NAVIO HANDPIECE; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BLUE BELT TECHNOLOGIES NAVIO HANDPIECE; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number PFSR110137
Medical Device Problem Code Circuit Failure (1089)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/19/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, during the preparation for a navio assisted tka surgery, they got an error message during bur homing saying "internal cabling drill/console error".They unplugged the navio handpiece and rebooted the machine.After start up, they got an error saying "pfs control hardware failure".The procedure was completed with manual instrumentation without significant delays.The patient was not harmed beyond the reported problem.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Additional Manufacturer Narrative
H3, h6: the navio handpiece, part number pfsr110137, sn(b)(6), used for treatment, was returned for evaluation.A relationship between the reported event and the device was established.Nothing was identified visually that contributed to the reported problem.A functional evaluation was performed.The evaluation followed performance verification pv0004, rev c.The reported problem was confirmed.The handpiece would not perform the test & produced a homing error.The gear would not turn by hand.The max t1 torque was 98.No additional functional non-conformances were identified.The most likely cause of this event is mechanical failure related to the motor.The surgical technique guide provides a ¿recovery procedure guidelines¿ table that provides guidelines for recovering to a fully manual procedure.A review of manufacturing and records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints found similar events.A historical escalation event review was completed.The review determined that prior escalation actions are applicable to the scope of this complaint, however, no further escalation action is required.The failure mode and associated risk have been anticipated within the risk file and the documented risk level is still adequate.As a part of corrective action a design change to help reduce build up within the motor.The failure mode will continue to be monitored through complaint investigation and trended through post market surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVIO HANDPIECE
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13480846
Report Number3010266064-2022-00086
Device Sequence Number5364179
Product Code OLO
UDI-Device Identifier00885556628515
UDI-Public00885556628515
Combination Product (Y/N)N
Initial Reporter StateHI
Initial Reporter CountryUS
PMA/510(K) Number
K191223
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPFSR110137
Device Catalogue NumberPFSR110137
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/03/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/19/2022
Supplement Date Received by Manufacturer07/15/2022
Initial Report FDA Received Date02/08/2022
Supplement Report FDA Received Date07/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-