| Model Number |
283889 |
| Medical Device Problem Code |
Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Udi: (b)(4).To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported by the customer in (b)(6) that during a rotator cuff repair surgery on (b)(6) 2022, it was observed that the barrel tornado burr 5.5mm blade device did not fit into the handpieces.It was unknown how the procedure was completed.There was a delay of five minutes to complete the procedure.There were no adverse patient consequences reported.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Udi: (b)(4).To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported by the customer in (b)(6) that during a rotator cuff repair surgery on (b)(6) 2022, it was observed that the barrel tornado burr 5.5mm blade device did not fit into the handpieces.It was unknown how the procedure was completed.There was a delay of five minutes to complete the procedure.There were no adverse patient consequences reported.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.D9, h3, h6: the actual device has been returned and is currently pending evaluation.Once the device has been evaluated, a supplemental medwatch report will be sent accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: h4: the device manufacture date was reported as unknown on the initial report; and has been updated accordingly.Investigation summary : device reported in this complaint was received for evaluation, a visual inspection and functional test were conducted.The result for this test indicated that the device could not be assembled into micro tornado handpiece (p/n 283512).The device was then measured, and it was identified that 2 burr features of the outer blade hub were over the upper limit; this information was sent to r&d lab confirmed that 2 burr features were oversized (12.575mm feature and 15.72mm feature).A manufacturing record evaluation was performed for the finished device lot number: m2106005, and no non conformances were identified.Manufacturing was notified about the condition and a nonconformance record was created to continue the investigation, root cause determination and mitigation of the issue.The supplier has conducted prior investigation indicating a probable root cause is unforeseen growth in outer blade hub (part that is assembled onto micro tornado handpiece) dimension post-induction welding process leading to oversized condition--particularly on hub components on higher end of specification pre-weld.Short term mitigations in place under a nonconformance record to implement additional inspection step post-induction welding process.The dimensional issue would result in difficulty of assembling the burr into the handpiece.Insertion, however, is still possible but will necessitate an extra amount of force as noted on the test performed.Full engagement is a requisite for optimal function.Once achieved, the device functions as intended with no issue.As during the investigation other lots were found to be affected, the failure mod was escalated to upper management.Depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
|
| |
|
Search Alerts/Recalls
|