| Model Number |
15004-095-090 |
| Medical Device Problem Codes |
Fracture (1260); Migration (4003)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
12/01/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device has not been returned for evaluation and a lot number was not provided.The root cause is unable to be determined at this time.If any relevant, additional information is provided, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2021, a (b)(6) y/o male underwent a spinal fusion at l2-l3 & l4-l5 ltp with t10-p psf.Around 2 months postoperatively, the patient reports to hearing a loud crack while getting up from a chair.The patient reports to having an increase in symptoms.A radiograph reveals screw loosening with broken iliac screw.A revision surgery was performed with l5-s1 tlif, replacement of l4 screws and connecting to superior construct.The broken shank remains in the patient.
|
| |
|
Search Alerts/Recalls
|