STRYKER ORTHOPAEDICS-MAHWAH DELTA V-40 CERAMIC HEAD 32/+4; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Model Number 6570-0-232
Medical Device Problem Code Inaccurate Information (4051)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/10/2022
Type of Reportable Event Malfunction
Event or Problem Description
Distributor reported that outer packaging marking indicated 32/+4 (6570-0-232) as did the inner packaging, however a 32/-4 (6570-0-032) was packaged.
 
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.Review of the device history records indicate devices were manufactured and accepted into final stock with no reported discrepancies.There have been 3 other events for the lot referenced.The following devices were also listed in this report: delta v-40 ceramic head 32/-4; cat#: 6570-0-032; lot#: unknown.It cannot be determined which, if any of these devices may have caused or contributed to the patient's experience.Not returned.
 
Event or Problem Description
Distributor reported that outer packaging marking indicated 32/+4 (6570-0-232) as did the inner packaging, however a 32/-4 (6570-0-032) was packaged.
 
Additional Manufacturer Narrative
The following devices were also listed in this report: delta v-40 ceramic head 32/-4; cat# 6570-0-032; lot# 89648802.It cannot be determined which, if any of these devices may have caused or contributed to the patient's experience.Reported event: an event regarding product mix involving a ceramic head was reported.The event was confirmed by review of the photographs provided.Method & results: product evaluation and results: the outer box of the device shows catalog number 6570-0-232, lot 89546202 on the label and also shows that the device was intended to be a 32mm head with a nk lnth of +4mm.A photo showing that the device packaged bears the marking 32mm -4.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been 8 other similar events for the lot referenced.Conclusions: the reported product mix was confirmed.Nc was raised to further address this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DELTA V-40 CERAMIC HEAD 32/+4
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
EI   NA
Manufacturer Contact
marisol santiago
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key13481261
Report Number0002249697-2022-00201
Device Sequence Number5388154
Product Code JDI
UDI-Device Identifier04546540608536
UDI-Public04546540608536
Combination Product (Y/N)N
Initial Reporter CountryEI
PMA/510(K) Number
K173499
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Distributor
Initial Reporter Occupation Other
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 04/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number6570-0-232
Device Catalogue Number6570-0-232
Device Lot Number89546202
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/11/2022
Supplement Date Received by Manufacturer03/08/2022
Initial Report FDA Received Date02/08/2022
Supplement Report FDA Received Date04/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/21/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
PFA 2902313
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age49 YR
Patient SexMale
-
-