| Model Number |
RA002-4545SLR |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
01/25/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device has not been returned for evaluation; however device is anticipated to be returned to (b)(4) lab.No conclusions can be drawn at this time.If any additional information is provided, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
Information was received that after five years of being implanted, a revision procedure was performed due to the rods only distracting 3 centimeters.After removal, the rod would not distract while utilizing manual distraction, however, a "click" could be heard.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Device records review: a review of device history records (dhr) confirmed the device met all quality inspections and specifications prior to release.
|
| |
|
Event or Problem Description
|
|
It was additionally reported that the surgeon believes he found a recurrent radiological symptom.During a distraction clinic the surgeon found that there is a visible asymmetry and when trying to distract the rods they failed to do so.
|
| |
|