| Model Number |
2012-03-030 |
| Medical Device Problem Codes |
Break (1069); Device-Device Incompatibility (2919); Scratched Material (3020)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/27/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Instruments are broken.No surgical delay.No adverse affects on patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information, which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained, that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected h6: (device code).
|
| |
|
Event or Problem Description
|
|
Additional information was received, stating that the bps broach sz 3, can't be put in a broach handle properly.And pin trl +4 ring is missing to hold the screw in the middle.Both the instruments, did not break into two or more pieces.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # : (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # = > (b)(4) investigation summary = > the device associated with this report was returned for analysis.Functional test did not confirm the complaint of unable to assemble.Broach assembled and disassembled properly with mating device.However, circular scratching is observed on the post of the broach.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot = > a worldwide lot specific complaint database search, or device history record (mre) review, was not possible because the required lot code was not visible.Corrected: h3.
|
| |
|