|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: the complaint device was received for evaluation.Upon reviewing the device, it could be observed that the inner shaft is seized, it was not possible to spin to any direction or separate.No anomalies were found on the black connector.As a potential cause cannot be associated to manufacturing, therefore a manufacturing record evaluation is not required.As part of depuy synthes mitek quality process all devices are manufactured, inspected, and released to approved specifications.According with the visual inspection, this complaint can be confirmed.A possible root cause for the reported issue can be attributed to the procedural variables, such handling of the device or product interaction during procedure; higher speeds applied on the handpiece can increase the temperature of the device's shaft, lack of adequate irrigation and excessive side loading can contribute to a seized inner shaft, as per ifu: adequate irrigation to the tip of the blade or burr is required to cool the blade and prevent accumulation of excised material in the surgical site.Excessive side loading may result in blade wear and degradation, as well as clogging and/or seizing.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to monitor additional complaint information for potential safety signals through complaint trending as part of post market surveillance.
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.D9, h3, h6: the actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|