SYNTHES GMBH CABLE CUTTER-LARGE; CUTTER, WIRE Back to Search Results
Catalog Number 391.906
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional narrative: reporter is a synthes employee.Part number: 391.906.Lot number: t940705.Manufacturing site: (b)(4).Release to warehouse date: (b)(6) 2009.A review of the device history records was performed for the finished device lot number, and no non-conformances were identified.Visual inspection: the cable cutter lrg was received at us customer quality (cq).Visual inspection of the complaint device showed the distal tip of the cutter had broken off, and one of the screws had come loose due to the stake breaking.Device failure/defect identified? yes dimensional inspection: a dimensional inspection was not performed due to post-manufacturing damage.Document/ specification review: current and manufactured revisions were reviewed.No design issues or discrepancies were identified.Complaint confirmed? yes.Investigation conclusion: this complaint is confirmed as the distal tip of the cutter had broken off, and one of the screws had come loose due to the stake breaking.No definitive root cause could be determined based on the provided information.There was no indication that a design or manufacturing issue contributed to the complaint.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that on an unknown date, during an inspection, the cable tensioner was discovered as defective.No specific issue was identified.No surgical impact.No patient involvement.This report is for one (1) cable cutter-large.This is report 2 of 2 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CABLE CUTTER-LARGE
Common Device Name
CUTTER, WIRE
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
SYNTHES TUTTLINGEN
unter hasslen 5
tuttlingen 78532
SZ   78532
Manufacturer Contact
kara ditty-bovard
1302 wright lane east
west chester, PA 19380
6107195000
MDR Report Key13483079
Report Number8030965-2022-00758
Device Sequence Number5363311
Product Code HXZ
UDI-Device Identifier07611819719880
UDI-Public(01)07611819719880
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number391.906
Device Lot NumberT940705
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/05/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/07/2022
Initial Report FDA Received Date02/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/08/2009
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CABLE TENSIONER.
Patient SexUnknown
-
-