EXACTECH, INC. LOGIC FEMORAL PS CEM LEFT SZ 2; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED Back to Search Results
Model Number LOGIC FEMORAL PS CEM LEFT SZ 2
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Osteolysis (2377)
Date of Event 11/10/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Concomitant medical products: logic tibia ps mod insert sz 2 13mm (cat# 02-012-35-2013 / serial# (b)(4)).Lgc tibial fit tray cem sz 2f / 2t (cat# 02-012-45-2020 / serial# (b)(4)).Three peg patella 32mm (cat# 200-02-32 / serial# (b)(4)).Additional information, including the product investigation, will be submitted within 30 days of receipt.
 
Event or Problem Description
As reported, approximately eight years post bilateral tka, this (b)(6) female patient had a grossly loose femoral component with severe osteolysis of distal femur and osteolysis of proximal tibial with partially loose tibial component.Patient had left tka revision requiring use of femoral augments and tibial cone.
 
Additional Manufacturer Narrative
After further review of additional information received the following sections g3, g6, h2, h3 and h6 have been updated accordingly.(h3 the revision reported was likely the result of osteolysis possibly related to wear of the tibial insert, which allowed for the femoral component and tibial tray to loosen from the bone.However, this cannot be confirmed because wear of the tibial insert was not reported, no images of the revised components were provided, and the devices were not returned to exactech for evaluation.The most probable root cause associated with the reported event of "osteolysis¿ is associated with destruction or disappearance of bone tissue likely related to weakened integration of an implant at the bone-implant interface.Furthermore, the most probable root caused associated with "prosthesis wear¿ is associated with material damage to a surface, usually involving progressive loss or displacement of material, due to relative motion between that surface and a contacting substance or substances.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOGIC FEMORAL PS CEM LEFT SZ 2
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer Contact
kate jacobson
2320 nw 66 court
gainesville, FL 32653
3523771140
MDR Report Key13483271
Report Number1038671-2022-00185
Device Sequence Number5462416
Product Code JWH
UDI-Device Identifier10885862001122
UDI-Public10885862001122
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K033883
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/30/2018
Device Model NumberLOGIC FEMORAL PS CEM LEFT SZ 2
Device Catalogue Number02-010-01-0220
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/19/2022
Supplement Date Received by Manufacturer01/13/2023
Initial Report FDA Received Date02/08/2022
Supplement Report FDA Received Date01/30/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/09/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
- LGC TIBIAL FIT TRAY CEM SZ 2F / 2T; - LOGIC FEMORAL PS CEM RIGHT SZ 2; - LOGIC TIBIA PS MOD INSRT SZ 2 13MM; - THREE PEG PATELLA 32MM
Outcome Attributed to Adverse Event Required Intervention;
Patient Age72 YR
Patient SexFemale
-
-