| Model Number |
9004000 |
| Medical Device Problem Code |
Sparking (2595)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/29/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
One capnograph was received for evaluation in relation to the reported event.Visual inspection of the device was able to confirm and replicate the reported event.The electric discharge that came out when the ac/dc adapter was plugged into the monitor was duplicated.The root cause of the failure has been determined to be due to the large capacitor on the main board of the power supply with role to accommodate the lack of smoothing caps on the output of the adapter.As a preventive action, the 9004 ifu was updated with the following statement: ?connect the ac charger to the ac power connector of the monitor first and the ac charger to the wall outlet second?.In addition, a warning label was added to the 1616xx power supply's output cable instructing the proper way to connect the power supply in order to eliminate spark.A dhr review was not performed as the issue is not related to a deficiency in the manufacture of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that a spark was emitted by this unit while being used by a patient.There was no patient injury reported.
|
| |
|