Model Number 82-6631
Medical Device Problem Code Connection Problem (2900)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/24/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The microsensor ((b)(4)) was not returned for evaluation (product discarded) therefore, an evaluation of the device could not be performed.Lot number information has been provided; therefore, device history record (dhr) was reviewed, and no anomalies were found.The root cause(s) of the reported issue could not be determined.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.The possible root cause of the defect reported by the customer could be due to excessive pressure applied to product.
 
Event or Problem Description
A facility reported the icp monitor could not detect the microsensor after being implanted; before implantation site closure.The surgeon changed with a new microsensor.The event led to 10 minutes surgical delay.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEUROMONITOR BASIC KIT
Common Device Name
ICP MICROSENSORS
Manufacturer (Section D)
INTEGRA LIFESCIENCES SWITZERLAND SAR
rue girardet 29
rue girardet 29
le locle
SZ 
Manufacturer (Section G)
INTEGRA LIFESCIENCES SWITZERLAND SAR
rue girardet 29
ch-2400
le locle
SZ  
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key13485362
Report Number3013886523-2022-00038
Device Sequence Number20104867
Product Code GWM
UDI-Device Identifier20886704042317
UDI-Public20886704042317
Combination Product (Y/N)N
PMA/510(K) Number
K914479
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 02/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2024
Device Model Number82-6631
Device Catalogue Number826631
Device Lot Number3713453
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/24/2022
Initial Report FDA Received Date02/08/2022
Date Device Manufactured05/08/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ICP MONITOR 826635
Patient SexUnknown
-
-