BOSTON SCIENTIFIC CORPORATION WALLSTENT ENDOPROSTHESIS; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number 26280
Medical Device Problem Codes Difficult or Delayed Positioning (1157); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/29/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device evaluated by mfr.: a visual examination of the returned device found that the stent had been fully deployed from the device.The delivery system was not returned for analysis.A visual examination found the distal wires of the deployed stent to be damaged.No other issues were identified with the device during analysis.
 
Event or Problem Description
Reportable based on device analysis completed on 17jan2022.It was reported that stent deployment failure occurred.The patient had thrombosis on both lower limbs and needed 5 wallstent stents to be implanted.A 75cm wallstent endoprosthesis self-expanding stent was advanced for treatment as a second stent; however, it could not be deployed.The device was removed and the procedure was completed with another of the same device.There were no complications nor injuries reported and the patient status was stable.However, returned device analysis revealed that the stent was detached and the stent wires were damaged.
 
Event or Problem Description
Reportable based on device analysis completed on 17jan2022.It was reported that stent deployment failure occurred.The patient had thrombosis on both lower limbs and needed 5 wallstent stents to be implanted.A 75cm wallstent endoprosthesis self-expanding stent was advanced for treatment as a second stent; however, it could not be deployed.The device was removed and the procedure was completed with another of the same device.There were no complications nor injuries reported and the patient status was stable.However, returned device analysis revealed that the stent was detached and the stent wires were damaged.It was further reported that the stent was intentionally deployed by the physician, after it was withdrawn from the patient, to see if the stent could be deployed outside the patient.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: a visual examination of the returned device found that the stent had been fully deployed from the device.The delivery system was not returned for analysis.A visual examination found the distal wires of the deployed stent to be damaged.No other issues were identified with the device during analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WALLSTENT ENDOPROSTHESIS
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13485429
Report Number2134265-2022-01089
Device Sequence Number5462003
Product Code MAF
UDI-Device Identifier08714729203971
UDI-Public08714729203971
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K152842
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/23/2022
Device Model Number26280
Device Catalogue Number26280
Device Lot Number0026237584
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/11/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/17/2022
Supplement Date Received by Manufacturer02/09/2022
Initial Report FDA Received Date02/08/2022
Supplement Report FDA Received Date02/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/23/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age63 YR
Patient SexMale
Patient Weight70 KG
-
-