APPLIED MEDICAL RESOURCES EA020, VOYANT GENERATOR, 1/BX; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number EA020
Medical Device Problem Code Energy Output Problem (1431)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 12/20/2021
Type of Reportable Event Serious Injury
Event or Problem Description
Procedure performed: sleeve gastrectomy.Event description" post-op oozing or bleeding incidents, the end-user claimed to be caused by the vessel sealing device, generator was removed as a precaution measure for further investigation.Additional information received on january 18, 2022: was there any bleeding during the surgery? no intra-operative bleeding.Is there any reason why the doctor associated the event to be related to the voyant generator as it only delivers energy? the or staff didn¿t keep the handpieces that were used in the surgeries, therefore, we took out the generator as a precaution measure for further investigation.Might the experience be related to the voyant handpiece that was used with it instead? we can¿t know, the hand pieces used in both cases were thrown away, therefore we need to check in case of any dysfunction in the generator itself.Is there any further background as to what occurred during the surgery? as informed they had no issues during the surgery, patients were delivered to er a week later with post-op bleeding.Additional information on january 31, 2022: yes, two patients were operated in the same day, suffered from severe intra abdomianal bleeding.First patient was found in shock state in the ward 8 hours post op.He was taken immediately to or and large amount of blood found intra peritoneal, about 3 liters most likely from omental side, staple line and ports were clear.First patient: hemostasis done ,blood and plasma transfusion then pt.Transferred to icu for 48 hrs.Then to ward for a week both patients are diabetics , they took clexane 40 mg preoperative as the routine protocol.Additional information received on february 3, 2022: the initial information were collected from the hospital, however, as per the surgeon who did performed both surgeries, the cases were done in (b)(6), first patient experienced blood loss at the same day 8 hours after the surgery, and the other the next day.Incidents were reported in (b)(6).Both surgeries were sleeve gastrectomy: first patient: 3 liters of blood found intra peritoneal.Patient has been discharged.Type if intervention: re-operated: hemostasis done, blood and plasma transfusion then pt.Transferred to icu for 48 hrs.Then to ward for a week.Patient status: patient has been discharged.
 
Additional Manufacturer Narrative
The event unit is anticipated to return to applied medical for evaluation.A follow-up report will be submitted upon completion of investigation.
 
Additional Manufacturer Narrative
The generator unit was returned to applied medical for evaluation, without the handpiece that was used in conjunction with the generator during the event.Testing was performed on the generator unit.However, the complainant's experience of insufficient hemostasis could not be confirmed as the unit functioned as intended.Applied medical had reviewed the details surrounding the event and related products and is unable to determine the cause of the reported event based on the evaluation of the returned unit.The probability and criticality of harm resulting from this failure have been evaluated and were found to be at an acceptable level.
 
Event or Problem Description
Procedure performed: sleeve gastrectomy.Event description: post-op oozing or bleeding incidents, the end-user claimed to be caused by the vessel sealing device, generator was removed as a precaution measure for further investigation.Additional information received on january 18, 2022: was there any bleeding during the surgery? no intra-operative bleeding.Is there any reason why the doctor associated the event to be related to the voyant generator as it only delivers energy? the or staff didn¿t keep the handpieces that were used in the surgeries, therefore, we took out the generator as a precaution measure for further investigation.Might the experience be related to the voyant handpiece that was used with it instead? we can¿t know, the hand pieces used in both cases were thrown away, therefore we need to check in case of any dysfunction in the generator itself.Is there any further background as to what occurred during the surgery? as informed they had no issues during the surgery, patients were delivered to er a week later with post-op bleeding.Additional information on january 31, 2022: yes, two patients were operated in the same day, suffered from severe intra abdomianal bleeding.First patient was found in shock state in the ward 8 hours post op.He was taken immediately to or and large amount of blood found intra peritoneal, about 3 liters most likely from omental side, staple line and ports were clear.First patient: hemostasis done ,blood and plasma transfusion then pt.Transferred to icu for 48 hrs.Then to ward for a week both patients are diabetics , they took clexane 40 mg preoperative as the routine protocol.Additional information received on february 3, 2022: the initial information were collected from the hospital, however, as per the surgeon who did performed both surgeries, the cases were done in december 20, first patient experienced blood loss at the same day 8 hours after the surgery, and the other the next day.Incidents were reported in (b)(6) 2021.Both surgeries were sleeve gastrectomy.First patient: 3 liters of blood found intra peritoneal.Patient has been discharged.Type of intervention: re-operated: hemostasis done, blood and plasma transfusion then pt.Transferred to icu for 48 hrs.Then to ward for a week.Patient status: patient has been discharged.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EA020, VOYANT GENERATOR, 1/BX
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
APPLIED MEDICAL RESOURCES
22872 avenida empresa
rancho santa margarita CA 92688
Manufacturer Contact
aaron fulcher
22872 avenida empresa
rancho santa margarita, CA 92688
9497135765
MDR Report Key13486210
Report Number2027111-2022-00352
Device Sequence Number14799531
Product Code GEI
UDI-Device Identifier00607915137601
UDI-Public(01)00607915137601(11)190909(30)01(10)1368361
Combination Product (Y/N)N
Initial Reporter CountrySA
PMA/510(K) Number
K182244
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberEA020
Device Catalogue Number101470971
Device Lot Number1368361
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/17/2022
Supplement Date Received by Manufacturer01/17/2022
Initial Report FDA Received Date02/08/2022
Supplement Report FDA Received Date03/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/09/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
VOYANT HANDPIECE.
Outcome Attributed to Adverse Event Hospitalization;
Patient SexUnknown
-
-