• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. VERSAJET II CONSOLE; LAVAGE, JET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. VERSAJET II CONSOLE; LAVAGE, JET Back to Search Results
Model Number 66800039
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/19/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that, the versajet ii console was broken, also console cannot be turned on due to fuse is blown.As this happened in a non-surgical environment, there was not patient involvement.
 
Additional Manufacturer Narrative
The information received by the manufacturer has been re-evaluated for mdr reporting and it was determined that this case does not meet the threshold for reporting and is a non-reportable event.Device failure to power up will not directly cause or contribute to serious injury or death, even in a clinical setting.Additional conditions must also exist (e.G.Surgical /treatment conditions; patient factors; the unavailability of back-up or alternate therapies for greater than 24 hours) before an injury could possibly occur, and even under those conditions, the risk of patient harm is low.This event is considered not reportable pursuant to 21 cfr part 803.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSAJET II CONSOLE
Common Device Name
LAVAGE, JET
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
Manufacturer (Section G)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK   HU3 2BN
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13486535
Report Number8043484-2022-00055
Device Sequence Number20136269
Product Code FQH
UDI-Device Identifier00040565124759
UDI-Public00040565124759
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K143115
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number66800039
Device Catalogue Number66800039
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/19/2022
Supplement Date Received by Manufacturer01/19/2022
Initial Report FDA Received Date02/08/2022
Supplement Report FDA Received Date03/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-