• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AESCULAP AG QUINTEX CONSTRAINED SCREW 4.5X17MM; SPINE SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AESCULAP AG QUINTEX CONSTRAINED SCREW 4.5X17MM; SPINE SURGERY Back to Search Results
Model Number SC493T
Medical Device Problem Codes Compatibility Problem (2960); Mechanical Jam (2983)
Health Effect - Clinical Codes Failure of Implant (1924); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/12/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Manufacturing site evaluation: investigation on-going.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted.
 
Event or Problem Description
It was reported that there was an issue with sc493t-quintex constrained screw 4.5x17mml.According to the complainant, the quintex screw was not locking into the plate.It also would not back out and a screw removal instrument was used to lift the plate out with the screw still in it.This occurred during a revision anterior cervical discectomy fusion (acdf) procedure.An additional medical intervention was required.The surgery was delayed by about 45 minutes.It was noted that the surgery was successful after a longer plate was implanted.Additional patient information is not available.The adverse event / malfunction is filed under aag reference (b)(4).Involved components: sc512t - 22 mm plate; sc403t - constrained screw 4/16mm.
 
Additional Manufacturer Narrative
Manufacturing site evaluation: investigation on-going.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted.
 
Event or Problem Description
It was reported that there was an issue with sc493t-quintex constrained screw 4.5x17mml.According to the complainant, the quintex screw was not locking into the plate.It also would not back out and a screw removal instrument was used to lift the plate out with the screw still in it.This occurred during a revision anterior cervical discectomy fusion (acdf) procedure.An additional medical intervention was required.The surgery was delayed by about 45 minutes.It was noted that the surgery was successful after a longer plate was implanted.Additional patient information is not available.The adverse event / malfunction is filed under aag reference (b)(4).Involved components: sc512t - 22 mm plate; sc403t - constrained screw 4/16mm.
 
Additional Manufacturer Narrative
Additional information/correction: b5 - type of event; lot number for involved component.D4 - lot number.D9 - product return date.D10 - involved component lot number.H3 - evaluation.H4 - manufacture date.H6 - codes updated.Investigation results: we made a visual inspection of the screw and the plate in their incoming (in- screwed) condition.Except slight wear marks we found no damages or deformation.We then took the removal tool provided for such cases and tried to unscrew the screw according to the instructions.The surgical technique brochure notes specific points to observe in the case.Device history review: the device quality and manufacturing history records have been checked for all leading device(s) lot numbers and the products found to be according to our specification valid at the time of production.There are currently no further complaints with this lot and error pattern at hand.The current failure rate is within the risk analysis and therefore acceptable as per procedures; severity was 2(5) and probability 2(5).Conclusion/preventive measures: the root cause for the problem is handling related.The screw could be removed easily and without any further damage following the steps explained in the instructions.Based upon the investigation results, a capa is not required.
 
Event or Problem Description
The adverse event is filed under aag reference (b)(4).Involved components: (b)(4) - quintex hybrid plate 1-level 22mm - lot 52638103.(b)(4) - quintex constrained screw 4.0x16mm - lot unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUINTEX CONSTRAINED SCREW 4.5X17MM
Common Device Name
SPINE SURGERY
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
Manufacturer (Section G)
AESCULAP AG
po box 40
tuttlingen, 78501
GM   78501
Manufacturer Contact
christian von der gruen
po box 40
tuttlingen, 78501
GM   78501
MDR Report Key13487095
Report Number9610612-2022-00013
Device Sequence Number20145395
Product Code KWQ
UDI-Device Identifier04046963814137
UDI-Public4046963814137
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K100243
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSC493T
Device Catalogue NumberSC493T
Device Lot Number52625707
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/12/2022
Supplement Date Received by Manufacturer10/12/2023
Initial Report FDA Received Date02/08/2022
Supplement Report FDA Received Date10/25/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/28/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SC403T; SC403T - LOT UNKNOWN; SC512T; SC512T - LOT 52638103
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-