| Model Number |
XIFOSS410 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
12/07/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the implant fractured resulting in it being removed.Tooth #19.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimmer biomet complaint number (b)(4).Patient weight not provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimmer biomet complaint number (b)(4).Patient weight not provided.
|
| |
|
Event or Problem Description
|
|
It was reported that the implant fractured resulting in it being removed.Tooth #19.
|
| |
|
Event or Problem Description
|
|
There is no update to the original complaint description provided.
|
| |
|
Additional Manufacturer Narrative
|
|
An osseotite® certain® 2 implant 4 x 10mm (xifoss410) was returned for investigation.Visual inspection of the as returned products identified worn markings due to usage, excessive bone/debris and a fracture on the collar.No pre-existing conditions were noted on the per.The reported device was located on tooth #19 and was used for approximately 4 years.Pictures or x-ray images were not provided.Review of appropriate documentation: documents reviewed: biomet 3i dental implant ifu (p-iis086gi) rev i - 2022/04/01.Dhr review: dhr review was completed for the subject lot number (2016021155).It was confirmed that all operations and inspections were executed as per applicable procedure.No deviations or non-conformances, which could have caused or contributed to the reported event was noted as part of the dhr.Lot was inspected and passed all acceptance criteria by qa.Complaint history review: complaint history review was performed for the reported lot number (2016021155) for similar event and no other complaint was identified.Review completed utilizing keywords: fracture : implant.Post market trend review: may post market trending was reviewed and there were no actionable events or corrective actions for the reported event or device.Based on the available information, device malfunction did occur and the reported event was confirmed.
|
| |
|
Search Alerts/Recalls
|