MALLINCKRODT MEDICAL SHILEY; TUBE TRACHEOSTOMY AND TUBE CUFF Back to Search Results
Model Number 5CN70A
Medical Device Problem Code Gas/Air Leak (2946)
Health Effect - Clinical Codes Dyspnea (1816); Unintended Extubation (4564)
Date of Event 01/16/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, the doctor came to replace the old unit when the patient's relative reported three things; leak from the ventilator, which normally had vt of 500-550, and with the post-insertion of the trach it dropped to vt of 200-250.In addition, the nurse has checked the pilot balloon, and it seemed the cuff was empty.Also, the patient experienced grunting noises (respiration).The patient's relative has inflated the cuff every 2-3 hours.The doctor has replaced the cannula with the same one (same lot), yet before inserting the trach, he has pre-tested the cuff, and it was working properly.3-4 hours later, the issue of leaks and symptoms appeared as if the cuff had been deflated again.The patient stated that with the new trach, there was difficulty breathing.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during use, the doctor came to replace the old unit when the family spotted three things following the replacement.One was a leak from the ventilator, which normally had vt of 500-550 where post-insertion of the trach dropped it to vt of 200-250.In addition, the nurse checked the pilot balloon, and it seemed the cuff was empty.Also, the patient experienced grunting noises (respiration).The family inflated the cuff every 2-3 hours.The doctor replaced the cannula with one of the same lot.Before inserting the trach, the doctor pre-tested the cuff to see if it was working properly, and so it did.And again, 3-4 hours later, the issue of leaks and symptoms appeared as if the cuff had been deflated again.The patient stated that with the trach, there was difficulty breathing.The device was replaced with a different device with no further issues.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHILEY
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
MALLINCKRODT MEDICAL
cornamaddy
athlone 3810
EI  3810
Manufacturer (Section G)
MALLINCKRODT MEDICAL
cornamaddy
athlone 3810
EI   3810
Manufacturer Contact
clay chandler
5920 longbow drive
boulder, CO 80301
3035306409
MDR Report Key13487396
Report Number8020889-2022-00033
Device Sequence Number20145527
Product Code JOH
UDI-Device Identifier10884521172463
UDI-Public10884521172463
Combination Product (Y/N)N
Initial Reporter CountryIS
PMA/510(K) Number
K142296
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/28/2025
Device Model Number5CN70A
Device Catalogue Number5CN70A
Device Lot Number20H0968JZX
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/23/2022
Supplement Date Received by Manufacturer02/09/2022
Initial Report FDA Received Date02/08/2022
Supplement Report FDA Received Date02/24/2022
Date Device Manufactured08/28/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age62 YR
Patient SexFemale
Patient Weight57 KG
-
-