| Catalog Number |
2C6700 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/11/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that an interlink straight type blood set was leaking from the injection site port.Blood was hung utilizing the blood set and it was observed that blood back flowed out of the injection site port.This issue was identified during patient use.To resolve the issue, a new set was utilized with no issues.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Device manufacturer address 1: (b)(4).The device has been received and the evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was added to h3, h4, h6 and h10.H10: the device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.All components were correctly placed and according to specifications.Functional testing was performed including clear passage and pressure testing; and the results were satisfactory; no leak was noted.Additional priming was performed and no issues noted.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|