• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ENDOVIVE SAFETY PEG KIT; TUBES, GASTROINTESTINAL (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ENDOVIVE SAFETY PEG KIT; TUBES, GASTROINTESTINAL (AND ACCESSORIES) Back to Search Results
Model Number M00566490
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event was approximated to (b)(6) 2022 as no event date was reported.(b)(4).The device has not been received for analysis.Upon receipt and completion of the problem analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation that an endovive safety peg kit push method was used during a percutaneous endoscopic gastrostomy placement procedure.The procedure date is unknown.During the procedure, the peg tube detached at transition zone while physician pulled it through the stoma site.It was reported that the detached portion was pulled out by the physician.The procedure was completed with a new endovive safety peg kit push method.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
Block b3 (date of event): date of event was approximated to (b)(6) 2022 as no event date was reported.Block h6 (device codes): device code a0501captures the reportable event of peg tube detached.Block h10: an endovive safety peg tube was returned.Visual assessment was performed on the joint where the peg tube meets the rigid tubing.No damage or defect was noted.Gentle pressure was applied to the joint by grasping the peg tube and rigid tubing in each hand and pulling them apart, but the peg tube would not detach from the joint.Additionally, the peg tube was noted to have been cut, and the remainder of the tube was not returned.Therefore, the reported event of peg detached was not confirmed.Based on the condition of the returned device, engineers determined that the peg tube was cut, which is part of the process for completing the procedure with the device.This indicates that the device may have been used to complete the procedure.Boston scientific has determined the most probable cause of this complaint is no problem detected.The correct complaint device was returned.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.A labeling review was performed and, from the information available, this device was used per the instructions for use (ifu) / product label.
 
Event or Problem Description
It was reported to boston scientific corporation that an endovive safety peg kit push method was used during a percutaneous endoscopic gastrostomy placement procedure.The procedure date is unknown.During the procedure, the peg tube detached at transition zone while physician pulled it through the stoma site.It was reported that the detached portion was pulled out by the physician.The procedure was completed with a new endovive safety peg kit push method.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOVIVE SAFETY PEG KIT
Common Device Name
TUBES, GASTROINTESTINAL (AND ACCESSORIES)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key13497108
Report Number3005099803-2022-00594
Device Sequence Number19951123
Product Code KNT
UDI-Device Identifier08714729748410
UDI-Public08714729748410
Combination Product (Y/N)Y
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K031538
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2022
Device Model NumberM00566490
Device Catalogue Number6649
Device Lot Number0028441930
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/20/2022
Supplement Date Received by Manufacturer03/01/2022
Initial Report FDA Received Date02/09/2022
Supplement Report FDA Received Date03/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/01/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-