Model Number G27023
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Insufficient Information (4580)
Type of Reportable Event Malfunction
Event or Problem Description
As reported, a performer introducer's tip broke during an unknown procedure.Additional information has been requested.
 
Additional Manufacturer Narrative
Pma/510(k) #: k171999.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Additional Manufacturer Narrative
Pma/510(k) #: k171999.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event or Problem Description
As reported, a performer introducer's tip broke during an unknown procedure.Additional information has been requested.
 
Additional Manufacturer Narrative
Summary of event: as reported, a performer introducer's tip broke during an unknown procedure.Investigation evaluation: reviews of the complaint history, device history record, instructions for use (ifu), manufacturing instructions, and quality control procedures were conducted during the investigation.The complainant did not return the complaint device to cook for investigation.Cook completed a review of the device history record (dhr).The dhr for the reported lot recorded no relevant non-conformances.A related lot recorded one possible relevant non-conformance for "distal end, difficult to wire, will not wire", however, this device was scrapped.A review of complaint history records shows no other related complaints associated with the complaint device lot.Cook has concluded that no nonconforming product from this lot exists in house or in the field.A document review was completed.Cook has concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.The instructions for use (ifu) provides the following information to the user related to the reported failure mode: how supplied ¿upon removal from package, inspect the product to ensure no damage has occurred.¿ cook has concluded that component failure unrelated to manufacturing contributed to this incident.The appropriate personnel have been notified and cook will continue to monitor for similar events.Per the quality engineering risk assessment, no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERFORMER INTRODUCER
Common Device Name
DYB INTRODUCER, CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key13497212
Report Number1820334-2022-00186
Device Sequence Number4097232
Product Code DYB
UDI-Device Identifier00827002270237
UDI-Public(01)00827002270237(17)241108(10)14336087
Combination Product (Y/N)N
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/08/2024
Device Model NumberG27023
Device Catalogue NumberRCFW-16.0P-38-35-RB
Device Lot Number14336087
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/27/2022
Supplement Date Received by Manufacturer05/18/2022
Initial Report FDA Received Date02/09/2022
Supplement Report FDA Received Date05/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/08/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-