| Model Number |
IPN044386 |
| Medical Device Problem Code |
Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
Skin Inflammation/ Irritation (4545)
|
| Date of Event |
01/19/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported: the "patient experienced throat swelling and red mark after using product".Device removed and a different device was used.Additional information stated that patient had small abrasion from the device.Polysporin topical treatment used.No serious or permanent injury reported.
|
| |
|
Event or Problem Description
|
|
It was reported: the "patient experienced throat swelling and red mark after using product".Device removed and a different device was used.Additional information stated that patient had small abrasion from the device.Polysporin topical treatment used.No serious or permanent injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The sample was returned for evaluation.A visual exam was performed and no abnormalities were found.The product was assembled correctly.A device history record review was performed on the lot number reported and no relevant findings were identified.The manufacturing site reports that a 100% visual exam is conducted after the assembly process; therefore, any defects would be detected prior to release from the manufacturing facility.The complaint report states that the device was used for one and a half weeks on the patient.The ifu for this product states to replace every 24 hours or as needed to prevent accumulation of secretions.In addition, the trach vent is designed for single use only and must not be cleaned and reused.Reuse of the device creates a potential risk of serious injury or infection.Reprocessing of medical devices intended for single use only may result in degraded performance or a loss of functionality.The complaint could not be confirmed as there were no issues found with the returned device.
|
| |
|
Search Alerts/Recalls
|