MAUDE Adverse Event Report: OSCOR INC. GUIDESTAR¿ STEERABLE GUIDING SHEATH; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OSCOR INC. GUIDESTAR¿ STEERABLE GUIDING SHEATH; INTRODUCER, CATHETER Back to Search Results
Model Number D141101
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Conclusion not yet available, evaluation in process.A follow-up report will be submitted as soon as the investigation is complete.
 
Event or Problem Description
After heliostar catheter insertion, the physician slowly aspirated the sheath with a small syringe, when no more bubbles were visible he flushed it a bit, without applying much pressure.When doing this, there was sprinkling from the side of the valve of the sheath handle.The physician also reported that this has been a common occurrence lately.No patient harm reported.
 
Additional Manufacturer Narrative
B5: on (b)(6) 2022 the customer stated that the product was used for an afib - paroxysmal procedure.There was no delay in the procedure; however, the patient outcome is unknown.It is unknown what could have caused the sheath - air leak.One 13.5f guidestar steerable guiding sheath was received with the dilator.There were no other accessories.Blood was found on and inside the sheath and dilator.According to the event description summary, there was sprinkling from the side of the valve of the sheath handle.The valve was viewed under magnification and looked normal with no anomalies.The sides of the hub and cap were also viewed under magnification and there were no cracks or any other anomalies found.The sheath was manually leak tested with the syringe connected to the stopcock and the tip occluded.The sheath did not exhibit leakage when tested using this method.The sheath was then tested using the iso 11070 for liquid leakage test method.The device was occluded at the distal tip and pressurized to 38kpa and held at this pressure for 30 seconds and no leakage was observed from the hemostatic valve.The sheath was further pressurized beyond 300kpa and no leakage was observed from the stopcock, sideport tubing or through the handle.The device history records were reviewed to confirm that the device passed all applicable in-process and final inspections.Returned device analysis revealed the sheath was within manufacturing specifications.The sheath did not leak when tested manually.The sheath also met the iso leakage test method requirement and did not leak through the hemostatic valve, hub or handle as shown in the customer's video.The handle, hub and hub caps looked normal and intact when viewed under magnification.According to the dhr, the sheath passed all in-process and final inspection steps including visual, mechanical, dimensional and leak testing.No manufacturing defects were found.Per manufacturing procedure (b)(4).Per qa procedure (b)(4).Per ifu the steerable sheath must be thoroughly flushed with either saline or heparinized saline and free of air prior to use to avoid air embolism to the patient.Aspiration and flushing of the sheath should be performed frequently to help minimize the potential for air embolism.For injecting or aspirating through the sheath, use the sideport only with stopcock.Prior to infusion, remove all air using the sideport.No further follow-up is required.Based on the investigation, a capa is not required as findings did not identify a design, labeling or manufacturing non-conformity.In addition, there was no new failure mode identified and the risk remains acceptable.The event will be re-evaluated if additional information becomes available.Oscor will continue to monitor this event type and risk.Oscor inc.Is submitting the above report to comply with 21 cfr part 803, the medical device reporting regulation.The report is based upon information obtained by oscor inc.Which the company may not have been able to investigate or verify prior to the date the report was required by fda.This report does not constitute an admission that the device, oscor inc., or its employees, caused or contributed to the event described in this report.Nor does this report reflect a conclusion by fda, oscor inc., or its employees, that the report constitutes an admission that the device, oscor inc., or its employees caused or contributed to the event described in this report.
 
Event or Problem Description
On (b)(6) 2022 the customer stated that the product was used for an afib - paroxysmal procedure.There was no delay in the procedure; however, the patient outcome is unknown.It is unknown what could have caused the sheath - air leak.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GUIDESTAR¿ STEERABLE GUIDING SHEATH
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
OSCOR INC.
3816 desoto blvd.
palm harbor FL 34683 1816
Manufacturer (Section G)
OSCOR INC.
3816 desoto blvd.
palm harbor FL 34683 1816
Manufacturer Contact
doug myers
3816 desoto blvd.
palm harbor, FL 34683-1816
7279372511
MDR Report Key13498118
Report Number1035166-2022-00026
Device Sequence Number11790544
Product Code DYB
UDI-Device Identifier00885672009496
UDI-Public00885672009496
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K140406
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/01/2023
Device Model NumberD141101
Device Catalogue NumberD141101
Device Lot NumberC8-19826
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/03/2022
Supplement Date Received by Manufacturer06/22/2022
Initial Report FDA Received Date02/09/2022
Supplement Report FDA Received Date06/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/28/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-