• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN MEDICAL INC. DESIGN OPTIONS®; SPINAL ANESTHESIA KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN MEDICAL INC. DESIGN OPTIONS®; SPINAL ANESTHESIA KIT Back to Search Results
Model Number 560606
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 12/19/2021
Type of Reportable Event Malfunction
Event or Problem Description
As per medwatch number mw5106280; spinal needle broke while being used for spinal analgesia.The patient was a (b)(6) male that was scheduled to have a right total knee arthroplasty.Diagnosis is: primary osteoarthritis of both knees.The plan for anesthesia was a peripheral nerve block and a spinal block.The needle broke during the spinal block.A second surgery needed to be planned to remove the needle.There was some delay because a spinal surgeon at 2 different hospitals were consulted prior to the surgery.He ended up staying a day in the hospital for the spinal surgery to remove the needle.The patient was unable to have his knee surgery and it was rescheduled for a later date.
 
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
 
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).No sample was provided for evaluation.Based on the data from the investigation we are unable to determine the root cause of the reported incident.The reported defect was unable to be confirmed.  review of the discrepancy management system (dsms) database was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.  we will maintain this report for further references and continue to monitor other reports for similar occurrences.  if any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DESIGN OPTIONS®
Common Device Name
SPINAL ANESTHESIA KIT
Manufacturer (Section D)
B. BRAUN MEDICAL INC.
901 marcon blvd.
allentown PA 18109
Manufacturer Contact
jonathan severino
901 marcon blvd.
allentown, PA 18109
4847197287
MDR Report Key13498409
Report Number2523676-2022-00023
Device Sequence Number11790545
Product Code OFU
UDI-Device Identifier04046964320958
UDI-Public(01)04046964320958
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2023
Device Model Number560606
Device Catalogue Number560606
Device Lot Number0061782587
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/31/2022
Supplement Date Received by Manufacturer01/31/2022
Initial Report FDA Received Date02/09/2022
Supplement Report FDA Received Date04/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/24/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age76 YR
Patient SexMale
-
-