As reported, the outer sheath of an 8mm x 150mm 120cm s.M.A.R.T.Sfa self-expanding stent (ses) was unable to move.As a result, the s.M.A.R.T.Stent was unable to be deployed.There was no injury to the patient.Additional information was requested but is not available and the device will be returned for evaluation.Per visual analysis, the stent of the unit was observed deployed approximately 5 mm.
|
|
The outer sheath of an 8mm x 150mm 120cm s.M.A.R.T.Sfa self-expanding stent (ses) was unable to move.As a result, the s.M.A.R.T.Stent was unable to be deployed.Per visual analysis, the stent of the unit was observed deployed approximately 5 mm.There was no injury to the patient.The device was returned for analysis.One non-sterile smart 120 150, sfa - 8x150mm was received for analysis inside a plastic bag.No original packaging was returned.The valve of the unit was received partially opened.Per visual analysis, the stent of the unit was observed deployed approximately 5 mm.The strain relief was observed bent as received.Also, the hypo tube rod of the unit was observed severely bent.No other anomalies were observed.The stroke length and usable length of the stent delivery system couldn¿t be measured due to the bent condition of the hypo tube rod and the strain relief of the unit, as received.Per functional analysis, deployment test of the stent was performed successfully; neither resistance nor any anomalies were observed during the test.No damages were observed on the stent.A product history record (phr) review of lot 18034532 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.The reported ¿stent delivery system (sds)-ses - deployment difficulty - partial deployment¿ was confirmed since the unit was received partially deployed.However, the unit was successfully deployed during deployment test, neither resistance nor any anomalies were observed during the test.The cause of the reported event and the condition of the unit, as received, could not be conclusively determined.However, it is probable that procedural and or handling factors such as vessel characteristics and the user¿s interaction with the device may have led to the reported event.According to the instructions for use (ifu) ¿advance the device over the guidewire and through the introducer to the target site.If resistance is met during delivery system introduction, the system should be withdrawn and another system used.If resistance is felt during retraction of the outer sheath do not force deployment.Carefully withdraw the stent system without deploying the stent.¿ neither the phr review nor the product analysis suggests that the reported event could be related to the manufacturing process of the unit.Therefore, no corrective or preventive actions will be taken at this time.
|