PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD Back to Search Results
Model Number 400SMTHXSFT0203
Medical Device Problem Code Difficult or Delayed Separation (4044)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/29/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has been returned and the investigation results are pending.A follow up mdr will be submitted upon completion of the device investigation.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the paraclinoid internal carotid artery (ica) using penumbra smart coils (smart coils), a penumbra smart coil detachment handle (handle), a non-penumbra microcatheter, a non-penumbra catheter, and a guidewire.During the procedure, one non-penumbra coil was successfully implanted.Next, a smart coil was successfully implanted using the handle.After advancing the next smart coil into the target location, the physician attempted to detach the coil with the handle; however, the coil did not detach.It was reported that the smart coil remained in a straight shape when the pusher assembly was retracted into the microcatheter.The smart coil unintentionally detached within the microcatheter; therefore, the physician used the guidewire to push the smart coil into the target location.During fluoroscopy, the physician identified a radiodensity object behind the proximal end of the smart coil but decided to implant it together with the smart coil using the guidewire.The identity of the object was not confirmed, and no attempts were made to remove the object from the patient.The procedure was completed using a non-penumbra coil and the same microcatheter.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
Evaluation of the returned smart coil confirmed the embolization coil was detached.Further evaluation revealed that the pet lock was separated, and the pull wire was retracted on the proximal end of the pusher assembly.This typically occurs during a successful detachment attempt and likely contributed to the embolization coil detaching during the procedure.Further evaluation revealed pusher assembly kinks, and the clear braided section was exposed on the distal flexible shaft.This damage may have a result of the difficulty detaching the coil experienced during the procedure.Evaluation was unable to confirm the origin of the reported translucent material.Penumbra products are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.Section h.Box 6.Conclusions code 1: 4316 - the investigation findings do not lead to a clear conclusion about the root cause of the reported issue.H3 other text: placeholder.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SMART COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key13498509
Report Number3005168196-2022-00038
Device Sequence Number20028489
Product Code HCG
UDI-Device Identifier00814548015705
UDI-Public00814548015705
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K160832
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/23/2022
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/01/2023
Device Model Number400SMTHXSFT0203
Device Catalogue Number400SMTHXSFT0203
Device Lot NumberF84411
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/20/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/13/2022
Supplement Date Received by Manufacturer02/25/2022
Initial Report FDA Received Date02/09/2022
Supplement Report FDA Received Date03/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/02/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
Date Report Sent to FDA01/01/2005
-
-