• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. FEMUR TRABECULAR METAL POSTERIOR STABILIZED (PS) STANDARD POROUS; PROTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. FEMUR TRABECULAR METAL POSTERIOR STABILIZED (PS) STANDARD POROUS; PROTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Adhesion(s) (1695); Pain (1994); Loss of Range of Motion (2032); Scar Tissue (2060); Swelling/ Edema (4577)
Date of Event 12/15/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).It was reported a patient underwent manipulation under anesthesia approximately one month after total knee arthroplasty.According to tabers medical dictionary ¿ manipulation of a joint is a mobilization technique, sometimes involving a rapid thrust or stretching of a joint, with or without anesthesia.With anesthesia is (mua).Per clinical orthropaedics and related research (how to treat the stiff total knee arthroplasty? a systematic review), mua is used to treat and resolve arthrofibrosis (scar tissue).This procedure is performed to increase articular motion and reduce chronic pain from arthrofibrosis.The indication for manipulation under anesthesia is related to limited range of motion and stiffness due to adhesions/scar tissue.Multiple mdr reports were filed for this event, please see associated reports: 3007963827-2022-00036, 0002648920-2022-00026.
 
Event or Problem Description
It was reported that the patient underwent an initial right total knee arthroplasty.Approximately one month later, the patient underwent manipulation under anesthesia for arthrofibrosis and stiffness.The implants remained intact.It was noted that the patient continues to have limited range of motion due to scar tissue/adhesions.Attempts have been made and no further information has been provided.
 
Event or Problem Description
It was reported that the patient underwent an initial right total knee arthroplasty.Approximately one month later, the patient underwent manipulation under anesthesia for arthrofibrosis and stiffness.The implants remained intact.It was noted that the patient continues to have limited range of motion due to scar tissue/adhesions.Attempts have been made and no further information has been provided.
 
Additional Manufacturer Narrative
(b)(4).It was reported a patient underwent manipulation under anesthesia approximately one month after total knee arthroplasty.According to tabers medical dictionary ¿ manipulation of a joint is a mobilization technique, sometimes involving a rapid thrust or stretching of a joint, with or without anesthesia.With anesthesia is (mua).Per clinical orthropaedics and related research (how to treat the stiff total knee arthroplasty? a systematic review), mua is used to treat and resolve arthrofibrosis (scar tissue).This procedure is performed to increase articular motion and reduce chronic pain from arthrofibrosis.The indication for manipulation under anesthesia is related to limited range of motion and stiffness due to adhesions/scar tissue.Multiple mdr reports were filed for this event, please see associated reports: 3007963827-2022-00036, 0002648920-2022-00026.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.
 
Event or Problem Description
It was reported that the patient underwent an initial right total knee arthroplasty.Approximately one month later, the patient underwent manipulation under anesthesia for arthrofibrosis and stiffness.The implants remained intact.It was noted that the patient continues to have limited range of motion due to scar tissue/adhesions.Subsequently, approximately thirteen months from initial implantation, a mri indicated tibial subsidence and tibial loosening.No further interventions have been reported.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Proposed component code: mechanical (g04) - femur.Reported event was confirmed by review of medical records.Medical records were provided and reviewed by a health care professional.Review of the available records identified the following: right knee aspiration, pain, swelling, and arthrofibrosis; right knee pain, arthrofibrosis, tibial subsidence, implant loosening; following mri: ambulates with antalgic gait, rom is 5 short of full extension to 95 degrees flexion.Device history record was reviewed and no discrepancies relevant to the reported event were found.Review of complaint history found no additional related issues for this item and the reported part and lot combination.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMUR TRABECULAR METAL POSTERIOR STABILIZED (PS) STANDARD POROUS
Common Device Name
PROTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key13499467
Report Number0001822565-2022-00358
Device Sequence Number17150165
Product Code MBH
UDI-Device Identifier00889024469907
UDI-Public(01)00889024469907(17)310607(10)65022787
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
K122745
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Study,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/31/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42500806402
Device Lot Number65022787
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/21/2022
Supplement Date Received by Manufacturer06/05/2024
12/30/2024
Initial Report FDA Received Date02/09/2022
Supplement Report FDA Received Date07/01/2024
12/31/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/14/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexMale
Patient Weight91 KG
Patient RaceWhite
-
-