| Model Number |
IPN028436 |
| Medical Device Problem Code |
Delivered as Unsterile Product (1421)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/08/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Intuitive surgical detected failed sterile barrier at incoming inspection.The failed sterile barrier appears (pictures) to be due to a poor heat seal.Intuitive surgical complaint number (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
|
| |
|
Additional Manufacturer Narrative
|
|
Qn#(b)(4).Per dhr the product 24 ea isi da vinci s 8mm obturator lot # 73a2100003 was manufactured on 01/04/2021 a total of (b)(4) pieces.Lot was released on 01/19/2021.Dhr investigation did not show issues related to complaint.From the pictures attached it was confirmed the defect reported by the customer "incomplete seal - sterility compromised".However it is necessary to receive the physical sample to perform a proper investigation, determine root cause & implement corrective actions.A verification of failure mode reported in the current manufacturing process was conducted as follows: 125 samples were taken from the current production p/n 420023 24 ea isi da vinci s 8mm obturator, lot # 73a2200754, the samples were visually inspected and issue reported " incomplete seal - sterility compromised" was not observed in the current manufacturing process.Revision of fmea-10-080 rev 03 was performed and the failure mode is already including it, no update is required.Failure mode "incomplete seal - sterility compromised" could be confirmed with picture attached, however it is necessary to receive the physical sample to perform a proper investigation, determine root cause & implement corrective actions.Failure mode "incomplete seal - sterility compromised" could be confirmed with picture attached.However it is necessary to receive the physical sample to perform a proper investigation, determinate the root cause & corrective actions.
|
| |
|
Event or Problem Description
|
|
Intuitive surgical detected failed sterile barrier at incoming inspection.The failed sterile barrier appears (pictures) to be due to a poor heat seal.Intuitive surgical complaint number (b)(4).
|
| |
|
Search Alerts/Recalls
|