| Model Number |
24674 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/24/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The 75% stenosed target lesion was located in the mildly tortuous and severely calcified iliac vein.A 12.0 x 80, 135cm mustang balloon catheter was advanced for dilatation.However, during second inflation at 8 atmospheres for 10 seconds, the balloon ruptured.The device was removed and the procedure was completed with a different device.There were no patient complications nor injuries reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Initial reporter address 1: (b)(6).
|
| |
|
Additional Manufacturer Narrative
|
|
Initial reporter address 1: (b)(6).
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The 75% stenosed target lesion was located in the mildly tortuous and severely calcified iliac vein.A 12.0 x 80, 135cm mustang balloon catheter was advanced for dilatation.However, during second inflation at 8 atmospheres for 10 seconds, the balloon ruptured.The device was removed and the procedure was completed with a different device.There were no patient complications nor injuries reported.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The 75% stenosed target lesion was located in the mildly tortuous and severely calcified iliac vein.A 12.0 x 80, 135cm mustang balloon catheter was advanced for dilatation.However, during second inflation at 8 atmospheres for 10 seconds, the balloon ruptured.The device was removed and the procedure was completed with a different device.There were no patient complications nor injuries reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: a mustang 12.0 x 80, 135cm, 14atm was received for analysis.A visual examination identified that the balloon was not folded which indicates that the device was subjected to positive pressure.A balloon pinhole was identified located approximately 31mm distal of the proximal markerband.An examination of the balloon material and markerbands identified no issues which could potentially have contributed to this complaint.The rated burst pressure for this device is 14 atm as per mustang specification.A visual examination of the markerbands identified no issues which may potentially have contributed to the complaint incident.A visual and tactile examination identified no kinks or damage to the shaft which may have potentially contributed to the complaint incident.A visual and tactile examination identified no damage to the tip which may have potentially contributed to the complaint incident.
|
| |
|