• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION MUSTANG; CATHETER, BILIARY, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION MUSTANG; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number 24674
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/24/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon rupture occurred.The 75% stenosed target lesion was located in the mildly tortuous and severely calcified iliac vein.A 12.0 x 80, 135cm mustang balloon catheter was advanced for dilatation.However, during second inflation at 8 atmospheres for 10 seconds, the balloon ruptured.The device was removed and the procedure was completed with a different device.There were no patient complications nor injuries reported.
 
Additional Manufacturer Narrative
Initial reporter address 1: (b)(6).
 
Additional Manufacturer Narrative
Initial reporter address 1: (b)(6).
 
Event or Problem Description
It was reported that balloon rupture occurred.The 75% stenosed target lesion was located in the mildly tortuous and severely calcified iliac vein.A 12.0 x 80, 135cm mustang balloon catheter was advanced for dilatation.However, during second inflation at 8 atmospheres for 10 seconds, the balloon ruptured.The device was removed and the procedure was completed with a different device.There were no patient complications nor injuries reported.
 
Event or Problem Description
It was reported that balloon rupture occurred.The 75% stenosed target lesion was located in the mildly tortuous and severely calcified iliac vein.A 12.0 x 80, 135cm mustang balloon catheter was advanced for dilatation.However, during second inflation at 8 atmospheres for 10 seconds, the balloon ruptured.The device was removed and the procedure was completed with a different device.There were no patient complications nor injuries reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: a mustang 12.0 x 80, 135cm, 14atm was received for analysis.A visual examination identified that the balloon was not folded which indicates that the device was subjected to positive pressure.A balloon pinhole was identified located approximately 31mm distal of the proximal markerband.An examination of the balloon material and markerbands identified no issues which could potentially have contributed to this complaint.The rated burst pressure for this device is 14 atm as per mustang specification.A visual examination of the markerbands identified no issues which may potentially have contributed to the complaint incident.A visual and tactile examination identified no kinks or damage to the shaft which may have potentially contributed to the complaint incident.A visual and tactile examination identified no damage to the tip which may have potentially contributed to the complaint incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MUSTANG
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13500446
Report Number2134265-2022-01307
Device Sequence Number5328534
Product Code FGE
UDI-Device Identifier08714729795285
UDI-Public08714729795285
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K141521
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/26/2024
Device Model Number24674
Device Catalogue Number24674
Device Lot Number0028097425
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/03/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/24/2022
Supplement Date Received by Manufacturer02/18/2022
Initial Report FDA Received Date02/09/2022
Supplement Report FDA Received Date03/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/27/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
-
-