| Model Number |
750170 |
| Medical Device Problem Code |
Inflation Problem (1310)
|
| Health Effect - Clinical Code |
Respiratory Insufficiency (4462)
|
| Date of Event |
08/24/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
One tracheostomy tube was received for evaluation in relation to the reported event.Sample was received without its original packaging and with their certificate of safe handling.Air cuff symmetry test was performed which found that the cuff inflated symmetry and the device passed inflation symmetry specifications.The reported event was unable to be confirmed or replicated.Root cause of the reported event was unable to be determined as the reported event was unable to be confirmed.A dhr review was performed subsequent to the manufacturing of the device and prior to its release.No problems or issues were identified during this dhr review.
|
| |
|
Event or Problem Description
|
|
It was reported that an asymmetric cuff was observed with the tracheostomy tube along with water being present.Issue occurred while in use with patient.Patient had respiratory difficulties.Airway was maintained by changing the tracheostomy which resolved the issue.
|
| |
|
Additional Manufacturer Narrative
|
|
This mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).A product sample was received for evaluation.Visual and functional testing were performed.Air cuff symmetry test was performed to the units by inflating with air and the cuff inflated completely.Based on the test, the units are within spec.The complaint is not confirmed.No root cause could be determined since the complaint was not confirmed.No actions taken were performed since the complaint was not confirmed.No problems or issues were identified during this device history record review.
|
| |
|
Search Alerts/Recalls
|