| Model Number |
ORT100 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/17/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The o-ring gasket at the ort100 table hydraulic pump/reservoir degraded over time and leaked.The gasket was replaced and the table was tested to verify proper function.This mdr is being submitted outside of the required timeframe as part of remedial action initiated by the manufacturer, in response to internally identified issues regarding failed electronic submissions through webtrader.
|
| |
|
Event or Problem Description
|
|
During a planned maintenance on (b)(6), 2019, an imris service engineer observed a small hydraulic fluid leak at the ort100 table pump/reservoir.The leak did not affect the table function and there was no patient involvement.
|
| |
|