• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. BIVONA MID-RANGE AIRE-CUF ADULT TRACH TUBES; TUBE TRACHEOSTOMY AND TUBE CUFF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. BIVONA MID-RANGE AIRE-CUF ADULT TRACH TUBES; TUBE TRACHEOSTOMY AND TUBE CUFF Back to Search Results
Model Number 750170
Medical Device Problem Code Inflation Problem (1310)
Health Effect - Clinical Code Respiratory Insufficiency (4462)
Date of Event 08/24/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
One tracheostomy tube was received for evaluation in relation to the reported event.Sample was received without its original packaging and with their certificate of safe handling.Air cuff symmetry test was performed which found that the cuff inflated symmetrically and the device passed inflation symmetry specifications.The reported event was unable to be confirmed or replicated.Root cause of the reported event was unable to be determined as the reported event was unable to be confirmed.A dhr review was performed subsequent to the manufacturing of the device and prior to its release.No problems or issues were identified during this dhr review.
 
Event or Problem Description
It was reported that an asymmetric cuff was observed with the tracheostomy tube along with water being present.Issue occurred while in use with patient.Patient had respiratory difficulties.Airway was maintained by changing the tracheostomy which resolved the issue.
 
Additional Manufacturer Narrative
This remediation mdr was generated under protocol b10010116, as a result of warning letter cms# 617147.Corrected data: b1: adverse event and/or product problem.H1: type of reportable event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIVONA MID-RANGE AIRE-CUF ADULT TRACH TUBES
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
5700 west 23rd ave
gary IN 46406
Manufacturer (Section G)
NULL
5700 west 23rd ave
gary IN 46406
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key13501530
Report Number3012307300-2022-03073
Device Sequence Number19812031
Product Code JOH
UDI-Device Identifier15021312006414
UDI-Public15021312006414
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K083641
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date09/01/2025
Device Model Number750170
Device Catalogue Number750170
Device Lot Number4033787
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/15/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/25/2021
Supplement Date Received by Manufacturer05/02/2023
Initial Report FDA Received Date02/09/2022
Supplement Report FDA Received Date05/10/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/28/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-