| Model Number |
133S-MX-15-T |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
01/06/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "patient that two deflated tissue expanders.¿ device remains implanted.Left side.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "patient that two deflated tissue expanders.¿ device remains implanted.Left side.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data:a.6, b.5, d.6b, h.3, h.6.
|
| |
|
Event or Problem Description
|
|
Device was explanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side "patient that two deflated tissue expanders.¿ device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: analysis of the returned device identified: opening on radius and yellow particles in the shell.Leak test and microscopic analysis was performed which identified: striated opening and broken on radius.The leak test is not performed because the shell is broken, based on the device analysis the final assessment is: striated opening and broken assessed as surgical damage.
|
| |
|