| Model Number |
9733857 |
| Medical Device Problem Code |
Imprecision (1307)
|
| Health Effect - Clinical Code |
Unspecified Tissue Injury (4559)
|
| Date of Event |
01/19/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A medtronic representative went to the site to test the equipment.The reported issue could not be confirmed or replicated, no components were replaced.The system then passed the system checkout and was found to be fully functional.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding a navigation system being used during a spinal procedure.During a procedure (l4-l5 fixation), after the initial image acquisition, a screw was inserted using both the navigation and a c-arm. after inserting four screws, once screw was determined to be misaligned in the c-arm image.It was confirmed all four placed screws were misaligned with a confirmation image acquisition from the imaging system.After removing the four screws, they were reinserted.The patient's vitals and other conditions did not change.The reported issue resulted in a procedure delay of less than one hour.The procedure was completed with continued use of the imaging system.There was no patient involvement.After the procedure, the accuracy was checked with the imaging and navigation systems.No migration could be confirmed in a test scan using the bone model.After the event, the procedure was performed and the behavior was not reproduced.There was no problem with device accuracy.The patient's bone was reported to be hard and it was hit with a hammer when the screw was implanted and the reference was misaligned.It was possible the error occurred and the screw was misaligned.The cause was also shared with the surgeon.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Additional information was received.The inaccuracy was not measured but believed to be around 6mm.The hospital staff accepted the rep's explanation that there might be a potential cause in the surgeon's procedure and further analysis was refused by them.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: a software analysis was initiated.However, the software evaluation found that a probable cause was unable to be determined.H6: fdm b01, fdr c19, and fdc d15 are applicable to the software analysis.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|