| Model Number |
KSP18140E |
| Medical Device Problem Code |
Loose or Intermittent Connection (1371)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
01/12/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient underwent a revision surgery on (b)(6) 2022 due to the boss of the resurfacing femur cracking.This fracture resulted in an incomplete surface to accept the morse taper of the stem extension resulting in the two implants dissociating.During the revision surgery, the following eleos implants were revised: resurfacing femur, resurfacing femur axial pin, tibial hinge component, and tibial poly spacer.No additional information regarding this adverse event has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is in process.When the investigation is complete, a supplemental mdr will be submitted accordingly.Multuple mdrs were submitted for this event: #3013450937-2022-00074.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted to include additional information.The investigation is complete.The device was not returned for evaluation.The root cause of the adverse event could not be determined.The device history records and sterilization batch records were reviewed and no issues during manufacturing or sterilization were identified that would have contributed to this complaint.If any additional information is obtained, a supplemental mdr will be filed accordingly.The following sections were updated: b4: date of this report added.G3: date received by manufacturer added.G6: type of report added.H2: follow-up type added.H3: device evaluated by manufacturer updated to no.H6: type of investigation code updated to 3331: analysis of production records.H6: type of investigation code updated to 4111: communication/interviews.H6: type of investigation code updated to 4110: trend analysis.H6: type of investigation code updated to 4114: device not returned.H6: type of investigation code updated to 4117: device not accessible for testing.H6: type of investigation code updated to 4109: historical data analysis.H6: investigation findings code updated to 213: no device problem found.H6: investigation conclusions code updated to 4315: cause not established.H10: additional narratives/data.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent a revision surgery on (b)(6) 2022 due to the boss of the resurfacing femur cracking.This fracture resulted in an incomplete surface to accept the morse taper of the stem extension resulting in the two implants dissociating.During the revision surgery, the following eleos implants were revised: resurfacing femur, resurfacing femur axial pin, tibial hinge component, and tibial poly spacer.No additional information regarding this adverse event has been provided.
|
| |
|