| Model Number |
66800045 |
| Medical Device Problem Codes |
Burst Container or Vessel (1074); Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/20/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, during the set up for a debridement treatment, the water hose of a versajet ii hydrosurgery system handset, 45deg 8mm plus burst at the transition to the handpiece.Treatment was performed, without any delay, with a back-up device instead.Patient was not injured as consequence of this problem.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: case (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: case (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that, during the set up for a debridement treatment, the water hose of a versajet ii hydrosurgery system handset, 45deg 8mm plus burst at the transition to the handpiece.Treatment was performed, without any delay, with a back-up device instead.Patient was not injured as consequence of this problem.
|
| |
|
Additional Manufacturer Narrative
|
|
Updated b5, h2, h3, h6.The investigation into the reported event is now complete, the complained device has not been returned for evaluation, the event cannot be confirmed.A documentation review, observed no historical, manufacturing or quality concerns.The reported event is adequately mitigated within the risk files, the instruction for use, (ifu), provides adequate delineation.The ifu calls out in set-up, "ensure there are no kinks or other external obstructions in saline supply, high-pressure and waste evacuation hoses".With no manufacturing problems observed, no corrective actions are deemed necessary.The probable root cause may include hose rupture due to a kink within the hose.Smith and nephew can confirm the device was released according to specifications and continue to monitor for adverse trends relating to this product range.
|
| |
|
Event or Problem Description
|
|
It was reported that, during the set up for a debridement treatment, the water hose of a versajet ii hydrosurgery system handset, 45deg 8mm plus burst at the transition to the handpiece.Treatment was performed, without any delay, with a back-up device instead.Patient was not injured as consequence of this problem.
|
| |
|