Catalog Number 383078
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/27/2022
Event Type  malfunction  
Manufacturer Narrative
A device evaluation and/or device history review is anticipated, but is not complete.Upon completion, a supplemental report will be filed.
 
Event Description
It was reported that bd closed iv catheter nexiva¿ contained foreign matter on the device catheter.The following information was provided by the initial reporter: "floccules foreign matter were found on the tip of catheter tube when the package was opened in pediatric surgery.".
 
Manufacturer Narrative
The following fields were updated due to additional information: d10: device available for eval yes, d10: returned to manufacturer on: 2022-02-10.H6: investigation summary a device history review was conducted for lot number 1293550.Our records show that this is the only instance of foreign matter occurring in this production batch.According to the sampling plan applied for product performance, this lot was accepted and released without defects being noted during the final assembly or visual inspections.Additionally, a sample was returned to our facility.Our engineers noted that the device was returned in an open packaging unit and that the device was contaminated with a thread-like foreign body.Microscopic inspection of the material was able to identify characteristics consistent with clothing fiber.The issue has been confirmed.Unfortunately due to the fact that that packaging was open, our engineers are not able to rule out on-site contamination, which has prevented them from confirming an association with the manufacturing process.H3 other text : see h10.
 
Event Description
It was reported that bd closed iv catheter nexiva¿ contained foreign matter on the device catheter.The following information was provided by the initial reporter "floccules foreign matter were found on the tip of catheter tube when the package was opened in pediatric surgery".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD CLOSED IV CATHETER NEXIVA¿
Type of Device
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BECTON DICKINSON MEDICAL DEVICES CO., LTD. SUZHOU
no.1 liangpu street
suzhou industrial park
suzhou 21512 6
CH  215126
Manufacturer (Section G)
BECTON DICKINSON MEDICAL DEVICES CO., LTD. SUZHOU
no.1 liangpu street
suzhou industrial park
suzhou 21512 6
CH   215126
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key13502048
MDR Text Key286664154
Report Number3014704491-2022-00063
Device Sequence Number1
Product Code FOZ
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/09/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number383078
Device Lot Number1293550
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/27/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/20/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-