• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. PROFORE LITE SYSTEM KIT CASE 8; BANDAGE, ELASTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. PROFORE LITE SYSTEM KIT CASE 8; BANDAGE, ELASTIC Back to Search Results
Model Number 66000415
Medical Device Problem Codes Device Difficult to Setup or Prepare (1487); Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/13/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned for evaluation, confirming the reported event.Visual inspection found some small holes in the coband layer that appear to be from excessive force used to unroll.Functional evaluation found that the roll was difficult to unwind with excessive force needed.A documentation review has been conducted, confirming previous complaints of this nature, with corrective actions assigned, which established and inadequate standard operating procedure as the root cause.The instructions for use contain comprehensive instructions on the safe operation and use of the device.The risk files mitigate the reported issue with no updates required.Smith and nephew can confirm the device was released according to specifications and continue to monitor for adverse trends relating to this product range.This investigation is now complete, with no additional corrective actions deemed necessary.
 
Event or Problem Description
It was reported that the 4th layer in a profore lite system kit case 8 is being too tight to pull apart.This is causing ripping as trying to apply and also there are presence of pin holes/tears in it.This was noticed during treatment set up causing a delay less than 30 min, ending up using a s+n back up dressing.Patient was not impacted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROFORE LITE SYSTEM KIT CASE 8
Common Device Name
BANDAGE, ELASTIC
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
Manufacturer (Section G)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK   HU3 2BN
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13502063
Report Number8043484-2022-00062
Device Sequence Number3992163
Product Code FQM
UDI-Device Identifier05031844002498
UDI-Public05031844002498
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 02/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number66000415
Device Catalogue Number66000415
Device Lot Number1370315
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/02/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/03/2022
Initial Report FDA Received Date02/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-