• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH 32MM +4 LFIT V40 HEAD; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH 32MM +4 LFIT V40 HEAD; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Model Number 6260-9-232
Medical Device Problem Codes Corroded (1131); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Foreign Body Reaction (1868); Discomfort (2330); Metal Related Pathology (4530)
Date of Event 01/18/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
 
Event or Problem Description
It was reported through the filing of a lawsuit that allegedly the patient was implanted with an lfit anatomic cocr v40 femoral head on his left hip on or about (b)(6) 2009 and was revised on (b)(6) 2021.It is further alleged that he suffered injuries as a result of implantation and explantation of the device(s) at issue, device recall and excessive levels of chromium and cobalt in his blood.
 
Additional Manufacturer Narrative
Reported event: an event regarding corrosion, altr and abnormal ion level involving a metal head was reported.The events were confirmed through clinician review of the provided medical records.Method & results: -device evaluation and results: not performed as product was not returned.-clinician review: a review of the provided medical records by a clinical consultant indicated: this inquiry reports the failure of a total hip replacement over 11 years after implantation due to trunnionosis and elevated ion levels.I can confirm that this event took place since i was able to review the revision operation report.Regarding the possible root cause of this event, i cannot determine it with certainty.Causes of trunnionosis are multifactorial including surgical technique factors, patient factors, and implant factors.-device history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the reported lot.  conclusion: a review of the provided medical records by a clinical consultant indicated: this inquiry reports the failure of a total hip replacement over 11 years after implantation due to trunnionosis and elevated ion levels.I can confirm that this event took place since i was able to review the revision operation report.Regarding the possible root cause of this event, i cannot determine it with certainty.Causes of trunnionosis are multifactorial including surgical technique factors, patient factors, and implant factors.No further investigation for this event is possible at this time.If additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
It was reported through the filing of a lawsuit that allegedly the patient was implanted with an lfit anatomic cocr v40 femoral head on his left hip on or about (b)(6), 2009 and was revised on (b)(6), 2021.It is further alleged that he suffered injuries as a result of implantation and explantation of the device(s) at issue, device recall and excessive levels of chromium and cobalt in his blood.Update based on medical review dated 26 march 2022.[.]"there was gray and tan colored synovium throughout the left hip consistent with a pseudotumor." [.] surgeon stated that the femoral head was challenging to remove.He found ¿substantial amount of corrosion over the trunnion as well as inside of the morse taper of the femoral head.This was consistent with trunnionosis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
32MM +4 LFIT V40 HEAD
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NJ NA
EI   NA
Manufacturer Contact
diana rogers
210 centennial avenue
centennial park, elstree
borehamwood WD6 3-SJ
UK   WD6 3SJ
2082386500
MDR Report Key13502313
Report Number0002249697-2022-00205
Device Sequence Number17418477
Product Code JDI
UDI-Device Identifier07613327013085
UDI-Public07613327013085
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K010757
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/30/2013
Device Model Number6260-9-232
Device Catalogue Number6260-9-232
Device Lot Number26256405
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/17/2022
Supplement Date Received by Manufacturer04/14/2022
Initial Report FDA Received Date02/09/2022
Supplement Report FDA Received Date05/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/21/2008
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexMale
-
-