Model Number H7493941820250
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/15/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that shaft break occurred.A 2.50 x 20mm synergy xd drug-eluting stent was selected for use.However, upon removal of the stent from the sterile loop, it was broken in half at the mid shaft.The stent was never used and the procedure was completed with another of the same device.There were no patient complications nor injuries reported.
 
Event or Problem Description
It was reported that shaft break occurred.A 2.50 x 20mm synergy xd drug-eluting stent was selected for use.However, upon removal of the stent from the sterile loop, it was broken in half at the mid shaft.The stent was never used and the procedure was completed with another of the same device.There were no patient complications nor injuries reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: synergy xd mr us 2.50 x 20mm stent delivery system was returned for analysis in 2 sections.A visual examination of the stent found no issues.There was no sign of damage, stretching or lifting of the stent struts.The stent showed no signs of movement and was set between the proximal and distal marker bands.The crimped stent outer diameter was measured and the result was within maximum crimped stent profile measurement.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed no signs of damage.A visual and tactile examination of the hypotube found multiple kinks and a hypotube break approximately 22.5cm distal to the distal strain relief.A visual examination of the outer and inner lumen and mid-shaft section found no issues along the shaft polymer extrusion.No other issues were identified during the product analysis.
 
Event or Problem Description
It was reported that shaft break occurred.A 2.50 x 20mm synergy xd drug-eluting stent was selected for use.However, upon removal of the stent from the sterile loop, it was broken in half at the mid shaft.The stent was never used and the procedure was completed with another of the same device.There were no patient complications nor injuries reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY XD
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13503404
Report Number2134265-2022-01092
Device Sequence Number4033893
Product Code NIQ
UDI-Device Identifier08714729980810
UDI-Public08714729980810
Combination Product (Y/N)Y
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/06/2023
Device Model NumberH7493941820250
Device Lot Number0027149429
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/19/2022
Supplement Date Received by Manufacturer03/25/2022
Initial Report FDA Received Date02/10/2022
Supplement Report FDA Received Date04/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/06/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-